*Troxler model 3450 and 3430/3440 Plus/Enhanced models are $549.00 due to software fee.
*Troxler model 3450 and 3430/3440 Plus/Enhanced models are $549.00 due to software fee.
The FERITSCOPE FMP30 measures the ferrite content in austenitic and duplex steel according to the magnetic induction method. All magnetisable structure sections are measured i. e., in addition to deltaferrite also strain-induced martensite, for example, or other ferritic phases. It is suited for measurements according to the BaslerStandard and according to DIN EN ISO 17655. Areas of application are onsite measurements, e.g. of austenitic platings as well as weld seams in stainless steel pipes, containers, boilers or other products made of austenitic or duplex steel. Duplex steel is used increasingly in the chemical and petrochemical industries, e.g., for boilers and pipelines. A ferrite deficit in the weld seam area leads to strength reduction, an excess ferrite content to a reduction in toughness and ductility. In particular when welding duplex steel, the ferrite content in the welding area can easily assume unfavourable values either due to unsuitable welding filler materials or through poor heat input or heat removal. Only an onsite measurement can provide the assurance that the processing did not change the optimum ferrite content in an unfavourable manner at the expense of mechanical or corrosion-resistance propeties.